5-5
A high-pressure steam sterilizer is acceptable only for the
immersion attachment.
The Ultrasound probe (biometry probe, pachymetry probe, and A-scan diagnostic
probe) can be disinfected by immersing the contact section of the probe tip in
suitable cleaning fluids.
Biometer / A-diagnostic probe/Pachy probe
● PERFEKTAN®TB (4% solution)
One clinically tested cleaning product would be:
PERFEKTAN®TB
(4% concentration and an exposure times of 30 min)
Antimicrobial Properties Perfektan® TB is:
- bactericidal ( incl. MRSA=multi resistant Staphylococcus aureus) -
tuberculocidal - fungicidal (C. albicans), levurocidal - virus inactivating incl.
HBV/HIV/BVDV/HCV/Papova - Polyoma/Adeno virus - sporicidal Thoroughly
rinse the probe with purified water and thoroughly dry it after cleaning.
Within 10 mm