EasyManuals Logo

Tomey UD-800 User Manual

Tomey UD-800
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #175 background imageLoading...
Page #175 background image
8-11
<EMISSIONS>
<IMM
UNITY
>
Table
1
ENCLOSURE PORT
Test
Standard
Compliance
HOME HEALTHCARE
ENVIRONMENT
Conducted Emissions
CISPR 11
Group 1 Class B
Radiated Emissions
CISPR 11
Group 1 Class B
Harmonic distortion
IEC 61000-3-2
Harmonic Class A
Voltage fluctuations and flicker
IEC 61000-3-3
Complies
Test
Standard
IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
8 kV contact
2
kV
4
kV
8 kV,
15
kV
air
Radiated RF EM fields
IEC 61000-4-3
10
V/m
80MHz 2.7GHz
80% AM at 1kHz
Proximity fields from RF
wireless
communications
equipment
IEC 61000-4-3
See Table 2
RATED
power
frequency
magnetic
fields
IEC 61000-4
-
8
30 A / m
50Hz or 60Hz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tomey UD-800 and is the answer not in the manual?

Tomey UD-800 Specifications

General IconGeneral
BrandTomey
ModelUD-800
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals