EasyManua.ls Logo

Tonghui TH2827 Series - Page 96

Tonghui TH2827 Series
131 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TH2827 Series Operation Manual Chapter 7 Command Reference
90
ABSolute
FUNCtion:DEV<n>:MODE PERCent
OFF
Where,
ABSolute Absolute value deviation display
PERCent Percent deviation display
OFF Real value display
Where, <n> is
Character 1 (49) is equal to the nominal value of primary parameter.
Or character 2 (50) is equal to the nominal value of the secondary parameter.
For example: WrtCmd (“FUNC:DEV1:MODE ABS”)
Query syntax: FUNCtion:DEV<n>:MODE?
Return format:
ABX
PERC <NL^END>
OFF
The FUNCtion:DEV<n>:REFerence<value> command is used to set the nominal
value of the deviation. The FUNCtion:DEV<n>:REFernece<value>? query
returns the current nominal value of the deviation.
Command syntax: FUNCtion:DEV<n>:REFerence<value>
Where,
<value> is NR1, NR2 or NR3 data format.
<n> is
Character 1 (49) is equal to the nominal value of primary parameter.
Or Character 2 (50) is equal to the nominal value of the secondary parameter.
For example: WrtCmd (“FUNC:DEV1:REF 10”)
Query syntax: FUNCtion:DEV<n>:REFerence?
Return format<NR3><NL^END>
The FUNCtion:DEV<n>:REFerence:FILL command is used to set the nominal
value of the deviation. This command directs the instrument to make a test and
then copies the results of the primary and the secondary parameters as the
nominal values of the deviation.
Command syntax: FUNCtion:DEV<n>:RERerence:FILL
Where,
Character 1 (49) or character 2 (50) is equal to the nominal values of the

Table of Contents