2. Measurement Operation
31
7
Select the type of the reference luminance colorimeter.
Press the [ROTATION] button and select the type of the reference luminance colorimeter
("*" is displayed), then press the [ENTER] button. In the case of a BM-7A, select
"COMMON"
8
Select the number of the data to be written.
Press the [ROTATION] button to display the number of the data to be written
[ROTATION] : Data for the next number is displayed.
[SHIFT] : Each time the button is pressed, the display changes as shown below.
REF SMP K
REF : This means "Reference". Measurement data of the reference
luminance colorimeter is written.
SMP : This means "Sample". Measurement data of the BM-7A to be
corrected is written.
K : This means Correction factor. A correction factor that is
calculated based on data measured using REF and SMP, and
automatically input.
It is possible to store up to 5 items of correction factor data.
9
Press the RUN/HOLD switch.
The message "Measuring" is displayed on the LCD, and measurement by the reference
luminance colorimeter starts. After measurement end, the following screen is displayed.
Request
In the case that each of the various communication settings is not correct,
data communication cannot be performed. Check whether setting has
been performed properly according to the Error display.
Meaning of Error Displays
「1.5 Error Display」
10
Switch dip switch 8 of the BM-7A to be corrected to the WRITE side to write data.
SELECT REF DEVICE
SR-3
REF11
ttention !!
Please set
the PROTECT/WRITE SW
to the PROTECT side.
Memo