12
Instruction Manual AD Series
7.2 Calibration - External (NTEP Certified Legal for Trade Models Only):
External calibration can be performed with an external calibration weight equal to the scale’s capacity.
6. Enter the Weighing Mode by pressing the F key. Select “

by pressing the T key.
7. Use a flathead screwdriver to gently remove the Calibration Seal Screw located in the rear of the scale.
8. Once the screw has been removed and the calibration
switch exposed, use a pen or another pointing device to
toggle the calibration switch to the right (which is the ON
position).
9. Upon toggling of the calibration switch, the balance will
go into Calibration Mode.
10. To start calibration, press the M key.
11. Select “Cal On by pressing “T” key.


