Error or Indicator
Cause
Explanation / Solution
- - - -
Below Zero
Re-zero the scale by pressing the 0
key.
- -
Taring is not allowed
Place a weight on the pan before taking a
tare.
- -
Re-zeroing is not allowed
Remove weight from the pan and clear
stored tare values
L
Pan error
Make sure the pan is properly seated on
the pan support.
H
Exceeded capacity
The scale has exceeded its weighing
capacity. Reduce the weight.
Err b
Pan not cleared on power-up
The pan was not empty while the scale was
initiating at startup. Clear the pan and
restart the scale.
Sapl LO (Parts Counting)
Average piece weight is too low
The average piece weight is lower than 3e.
Be sure that individual piece weight is
greater than 3e.
Sapl LO (Percent Weighing)
Sample reference weight is too
low
The sample reference weight is lower than
100d. Increase the sample weight.
The scale will not turn on
Possible power failure
Check the power AC adapter connection.
Weighing results are not accurate
Inaccurate calibration
Calibrate the scale.
Weighing results are not
stabilizing
Irregular environment
Eliminate drafts or vibrations.