Appendix C–6
TAKING A DRIFT TEST
With the gauge still in an area free of other nuclear gauges or
radioactive sources and on the reference standard block, press
1 at
the Drift Test menu shown above.
NOTE
If it has been less than four hours or more than eight
hours since the last stat test, then the gauge allows the
operator to exit the drift test. To exit and take another
stat test, press the
YES key at the prompt. To continue
with the drift test, press the
NO key at the prompt.
The gauge prompts the operator to place the gauge on the reference
standard block with the source rod in the SAFE (shielded) position.
Ensure that the gauge is positioned properly and press the
ENTER
key.
The gauge displays the progress of the drift test. After taking the
drift test, the gauge displays the average counts and indicates if the
test passed or failed. If it fails, see page C–3
for instructions. Press
the
ENTER key to return to the Drift Test menu, or press the
ESC key to return to the Ready screen.
REVIEWING THE DRIFT TEST
To review the results of the last drift test, press 2 at the Drift Test
menu (see page C–
5). The gauge displays the results of the last drift
test. To scroll through the data, use the arrow keys. Press the
ENTER key to return to the Drift Test menu, or press the ESC
key to return to the Ready screen.