MOISTURE OFFSET
The moisture offset (k) is determined by comparing the moisture of
a laboratory sample with the gauge moisture reading. With the
Model 3450, the operator can either allow the gauge to derive the
moisture offset or enter the moisture offset manually. To select a
stored moisture offset or to allow the gauge to derive the moisture
offset, see below. To manually enter the moisture offset, refer to
page 6–8.
Select a Stored Moisture Offset
To select a moisture offset already stored in the gauge, access the
Offset menu by pressing
OFFSET. Access the Moisture Offset
menu by pressing
2. Select the Change Offset menu by pressing
3.
Press 1 Stored Offset. The gauge displays the stored moisture
offsets. Use the number keys to select a moisture offset. The gauge
enables the moisture offset and returns to the Ready screen.
Moisture Offset
1- Stored Offset
2-
Gauge Derived
3-
Keypad Entry