EasyManua.ls Logo

Troxler RoadReader Plus 3450 - Soil Mode

Troxler RoadReader Plus 3450
201 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4–10
SOIL MODE
To measure the moisture and density of soil, soil-stone aggregate,
and similar materials, follow the steps below:
Select the Soil mode (see page 3–23).
If desired, enter, change, or enable the Proctor value (see page
3–19).
Prepare the test site as described on page 4–6.
Place the gauge on the test area.
Insert the source rod into the hole made by the drill rod or
source rod adapter. Use care when inserting the source rod. Try
not to disturb the soil around the hole.
Lower the source rod to the measurement depth. Release the
trigger. Gently tap down on the handle over the source rod to
lock the source rod into position. You should hear a click when
the source rod locks into position.
Gently slide the gauge to the right (towards the keypad) so the
source rod makes contact with the wall of the hole.
Press the START key.
In the Manual depth mode (see page 3–14), the gauge will
prompt for the source rod depth. Enter the source rod depth
using the number keys. For example, with the source rod in
the backscatter position, press
0. Then press ENTER.
In the Automatic depth mode, the gauge software reads the
depth strip on the source rod to determine the source rod
depth.
Note that the Proctor value is based on soil particles 3/8
inches or less in diameter (or those passing a #4 sieve). If an
unusual reading is obtained and oversize particles are
suspected, rotate the gauge 90°. Use the same drill hole to
take a second reading.

Table of Contents