Model 3450
6–3
WET DENSITY OFFSET
The wet density offset is the difference between the gauge reading
and the actual density. To access the wet density offset functions,
press
OFFSET to access the Offset menu. Then press 1 to access
the Wet Density Offset.
The gauge displays the current wet density offset on the second line.
To enable the offset, press
1. To disable the offset, press 2.
To enter a new wet density offset, press 3. The gauge prompts for
the wet density offset.
To change the offset, select the offset sign (positive or negative).
Enter the wet density offset. Press
ENTER. The gauge enables the
new wet density offset and returns to the Ready screen.
Wet Density Offset:
xxxx kg/m3
1-Enable 2-Disable
3-Change Offset
Wet Density Offset
xxxx kg/m3
Select
(+/-)