Flow Analyzer Overview 2-5
M e a s u r e m e n t s a n d U n i t s
Available Measurement Parameters
The list parameters available for selection is dependent on the model of the connected flow module.
Plateau pressure pressure at
the end of the inspiratory phase
during no flow (4081 only)
Peak flow rate Peak inhaled
flow rate
High pressure measurement
(4081 only)
Minimum flow rate the negative
of the peak exhaled flow rate in a
bi-directional test circuit.
Absolute pressure in flow tube.
Equals barometric pressure if
flow tube is open to atmosphere.
Flow rate delta the numerical
difference in flow rates.
Oxygen concentration
measurement (4081 with
4073 only)
Flow rate delta percent the
difference in flow rates by %.
Flow tube temperature
measurement.
Inhaled minute tidal volume
Inspiratory time including
pause time
Low pressure measurement
(4081 only)
Inspiratory rise time amount of
time from the start of breath to
the max flow rate
Peak inspiratory pressure
(4081 only)
Peak end expiratory pressure
(4081 only)
Inspiratory to Expiratory ratio
Mean airway pressure
(4081 only)
Inspiratory to Expiratory ratio
including pause time
Minimum low pressure
(4081 only)
Static compliance calculated
as Tidal volume / (plateau
pressure PEEP)
Delta low pressure Delta
airway pressure (4081 only)
Leak rate calculated as change
in low pressure over the set
Average Time