Page 33
Slow Exhaust: Steam is exhausted from the chamber at a slow
rate, until it reaches the pressure of 30 kPa above the ambient
pressure. Then steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Drying: N/A
Ending: The system checks that End temperature and End
pressure have been reached, and then releases the door.
5.5 Program 5: Liquid A – Cooling (C – cooling models
only)
For liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250F). for 15 minutes.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
Sterilization temperature:121C (250F).
Sterilization time: 15 minutes.
Operations sequence:
(PV, pre-vacuum models only)
Pulse low/Pulse high: at one
pulse it will build vacuum down to 25 kPa.
Heating: Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization: Sterilization temperature is maintained constant
during the sterilization time.