Page 78
Caution!
Do not touch the strainer’s cover, mounted on the exhaust
line, during and short after operation.
Touching the hot strainer’s cover may cause severe injuries.
10.7 Unloading
When the cycle ended, press START/STOP key to verify the
status. Message "Cycle Ended" (and the relevant failure
message, if applicable) is displayed on the screen.
1. Verify that there is no pressure in the chamber, according to the
reading on the display. Only then you may open the door.
2. Open the autoclave. (see sec. 10.2 Opening the door)
Warning!
To avoid severe injuries from hot steam when opening the
door:
It is strictly forbidden to lean on the autoclave.
It is strictly forbidden to place your hand or any part of your
body over the door.
Wear heat-resistant gloves or use the tray handle to remove
the load from the autoclave
To avoid severe injuries from hot steam when opening the
door:
It is strictly forbidden to lean on the autoclave.
It is strictly forbidden to place your hand or any part of your
body over the door.
Since the autoclave is defined as a bio-hazard autoclave, a
failed cycle may leave un-sterilized waste in the chamber.
Therefore failed cycle will be followed by a slow exhaust and
the door will remain locked. To enable opening the door
perform a new cycle. The door can be opened only after a
successful cycle.
Wear heat-resistant gloves or use the tray handle to remove the
load from the autoclave
On completion of the cycle, the load shall be visual inspected to
ascertain that the load is dry, and that the colour of the
sterilization indicators turned to the required colour.