Page 42
5.14 Program 14: Warm-Up
Pre-warming the chamber at 80C (176F) without drying.
Caution!
This is not a sterilization program!
Nominal parameters default settings
Warm-up temperature:
80
C (
176
F).
Warm-up time: 20 minutes.
Operations sequence:
Pulse low/Pulse high: N/A
Heating: Steam enters the chamber and heats it up until the
warm-up temperature is reached.
Warm up: Warm up temperature is maintained constant for the
warm-up time.
Sterilization: N/A.
Cooling: N/A.
Fast Exhaust: Steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Drying: N/A.
Note: Some water may remain in the chamber.