Page 36
Drying: vacuum is built up and the air pump is working.
5.8 Program 8: Hollow Load (PV – pre-vacuum models
only)
For class B sterilization of pipes and wrapped pipes, double wraps,
when the manufacturer recommends autoclaving at temperatures
of 121C (250F).
Nominal parameters default settings
Sterilization temperature: 121C (250F)
Sterilization time: 15 minutes
Drying time: 15 minutes.
Operations sequence:
Pulse low/Pulse high: four vacuum pulses remove air by
drawing the chamber down to 25 kpa.
Heating: Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization: Sterilization temperature is maintained constant
during the sterilization time.
Cooling: N/A
Fast Exhaust: the steam is exhausted out of the chamber at a
fast rate until pressure decreases to ambient pressure.
Drying: vacuum is built up and the air pump is working.