Page 81
11.Service and Maintenance
Instructions
11.1 Preventive and Scheduled Maintenance
The maintenance operations described in this chapter must be
fulfilled periodically in order to keep the autoclave in good working
condition and to reduce the breakdown time to a minimum.
The user maintenance personnel can easily execute these
operations in accordance with further instructions.
The owner of the autoclave is responsible to order an authorized
technician to perform the periodical tests and preventive
maintenance operations, as specified in the technician manual.
Use only mineral-free water as detailed in sec. 3.11 (water quality).
Warning
Before carrying out any preventive maintenance operation,
ensure that the electrical cord is disconnected and there is no
pressure in the autoclave.
11.2 Daily
Clean the door gasket with a soft cloth. The gasket should be
clean and smooth.
11.3 Weekly
1. Once a week clean and descale the chamber, copper tubes and
the reservoir.
2. Remove the baskets (if applicable). Clean the chamber and
baskets with a cleaning agent & water and with a cloth sponge.
Caution
Do not use steel wool or steel brush as this can damage the
chamber!
3. Put a few drops of oil on the two door pins and door tightening
bolts.
4. Clean outer parts of the autoclave with a soft cloth.