5.15 Program 15: Isothermal
For liquids when the manufacturer recommends autoclaving at 60-
100C (140-212F) with no drying. Recommending for melting of
agar, pasteurization etc.
Caution!
This is not a sterilization program!
Put one PT100 inside the bottle, leave the second one hanging in
the chamber outside the bottle (see below).
Nominal parameters default settings
Heating temperature (default): 80C (176F).
Heating time: 20 minutes.
Operations sequence:
Heating: Steam enters the chamber and heats it up until the
Keep Heat temperature is reached.
Keep Heat: Keep-Heat temperature is maintained constant for
the preset Keep-Heat time.