Page 47
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
Sterilization temperature: 121C (250F).
Sterilization time: 15 minutes.
Operations sequence:
(PV, pre-vacuum models only) Pulse low/Pulse high: at one
pulse it will build vacuum down to 25 kPa.
Heating: Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization: Sterilization temperature is maintained constant
during the sterilization time.
Slow Exhaust: Steam is exhausted from the chamber at a slow
rate, until chamber temperature reaches 105C.
Cooling: Forced cooling to the required end temperature,
chamber pressure remains at approx. 140kPa, and the cooling
valve is opened.
Fast Exhaust: The steam is exhausted out of the chamber at a
fast rate until pressure decreases to ambient pressure.
Drying: N/A
Ending: The system checks that End temperature and End
pressure have been reached, and then releases the door.