Page 63
Note: For all the standard sterilization cycles, and for Bowie and Dick
test, the only changeable cycle parameter is dry time (you will not
see other parameters on your screen). For the custom cycles
created by duplication, and for the Warm Up cycle, more options
are changeable. Operator is not allowed to create custom cycles;
only authorized technician can do this.
Dry Time
This parameter defines duration of the drying stage for the cycle.
1. Browse to the following folder:
Cycle parameters\Drying\Dry Time
2. Change the parameter as desired.
Drying stage is divided into two stages. For each stage you can
set the total time, on time and off time. So during the Dry First
Stage, the heating elements are on for the time set in Dry Heat
On 1, then they go off for the time set in Dry Heat Off 1, and then
this sequence is repeated during the entire Dry First Stage Time.
The same is true for the Dry Second Stage.
Add Dry Time
This subdirectory allows you to change the Dry Time parameter
for the current cycle.
Browse to the following folder:
Cycle parameters\Drying\Add Dry Time
Change the parameter as desired.