9.  Give the CALIBRATE CHECK command. The command takes 
about 90 seconds to complete. 
10.  The value produced by the check command must be close to the 
value printed on the scattering plates. If the difference is less than 
3%, the calibration is correct. In this case, leave the opaque glass 
plates in place, write down the signal value, and continue with 
section FSM102 Mechanical Alignment Check Procedure on page 
130. If the difference is more than 3%, leave the opaque glass 
plates in place and continue with the calibration procedure. 
 
  0201-077 
Figure 41  Assembling and Attaching Calibrator 
FSM102 Calibration Procedure 
The calibration procedure is performed as follows: 
1.  Follow the steps in section FSM102 Calibration Check Procedure 
on page 128 (if not yet done). 
2.  Install the calibrator assembly on the measurement unit (if not yet 
done).