Chapter 6 _______________________________________________________________ Maintenance 
 
VAISALA _______________________________________________________________________ 139 
Calibration procedure finished. 
Calibration: FAILED. 
Zero test: fail (3.7829, limit: 
absolute +/-0.0001) 
Signal test: none 
Remove calibrator assembly and type 
YES: YES 
 
1> 
Procedure failed due to Zero 
Signal Test failure. Zero 
Value is measured to 3.78. 
Limit is ±0.0001. 
Scatter Signal Test was 
skipped. 
 
Replacing FS11 Modules 
This section describes in detail how to remove and replace the Visibility 
Sensor Controller FSC102, transmitter module FST102, and receiver 
module FSR102. Removing the units comes into question when there is 
reason to suspect that a malfunction of FS11 is caused by faults in the 
optical units or the rain detector. 
CAUTION 
The servicing of the equipment must be performed by qualified 
personnel only.
 
 
CAUTION 
AC (mains) power switch and the optional battery backup switch 
must always be turned off before attempting any service actions 
described in this section.
 
 
NOTE 
When replacing any assembly, do not let water or moisture enter the 
unit.