Chapter 6 _______________________________________________________________ Maintenance 
 
VAISALA _______________________________________________________________________ 133 
8.  The instrument continues: 
Step 3/3 
 Remove zero plugs, install glass plates and enter 
signal value (between 0 and 10): 1.3497 
 Is signal value 1.349700 correct? 
 Type YES or NO? YES 
 1st run 
 ..............|||| 
 Signal test PASSED. (1.3514) 
Calibration procedure finished. 
Calibration: PASSED. 
Zero test: ok (0.0000, limit: absolute +/-0.0001) 
Signal test: ok (0.1259, limits: recalibration +/-
0.0300  rejection +/-0.2000) 
Remove calibrator assembly and type YES: YES 
1> 
 
9.  Remove the zero plugs from the measurement unit transmitter and 
receiver. 
10.  Install the calibrator assembly on the measurement unit as shown in 
Figure 41 on page 129. 
-  The assembly is attached to the measurement unit and the 
Scatter Signal Value is verified against the known scatter 
simulation value of the calibrator assembly (Signal Value given 
by the type label attached to the calibrator assembly). The test is 
passed successfully when the evaluated Scatter Signal Value 
deviates no more than ±3% to the known scatter simulation 
value. For deviations above ±3% and up to ±20%, the 
measurement unit calibration will be updated using the 
parameters of the calibrator assembly in use. The test has failed 
when the deviation is greater than ±20%. 
11.  Move away from the optical path and enter the signal value written 
on the calibrator assembly to the data terminal.