Page 101 of 244
2.6.2. CT Collimator Alignment with EzAlign
Follow the steps below to set CT collimator using EzAlign:
1. Start the EzAlign-S and click Modality Align.
2. Click CBCT and START. After clicking these buttons, the system moves to its capturing
position and the collimator auto-alignment starts.
If you want to use the auto irradiation function, click AUTO EXPOSE.