4.3.2. Before PANO sensor calibration
1. Remove the Normal Chin Block from the equipment.
2. Disassemble the Tube case (Refer to Chapter 1. Hardware Part Service Guide).
3. Rename the CAL folder (C:\VCaptureSW\Acquisition\PANO\WidePANO\CAL) name for backup.
If you are using an old collimator, ensure to reassemble the cu filter as illustrated below.
Before reassembling Cu filter After reassembling Cu filter
New Collimator Information is as listed below:
Application date: 2021-08-20
Cut in S/N: 098-000155