Page 133 of 244
4.3.3. PANO Sensor Setting
VAKCAP.exe file is required for PANO Sensor Calibration operation.
Path: C:\VCaptureSW\Acquisition\PANO\WidePANO
Before the PANO Sensor Calibration operation begins, the operator must
back up the prior PANO Sensor Calibration folder.
PANO Sensor Calibration file folder :
C:\VCaptureSW\Acquisition\PANO\WidePANO\CAL
To set the PANO sensor, use the command from the list below:
Command Function
[SPM_PANO] PANO mode
[SPM_PSON] PANO sensor power on
[SPM_FISS_0001] Normal Mode : 4x4 binning & High gain
[SPM_FIPM_0004] Sensor size setting
[SPM_FISM_0001] Sensor clock (1:External, 0: Internal)
[SPM_FREQ_0238] Sensor frequency 238 Hz
[SPM_COLM_0004]
1T Cu + AL filter position (4-Point)
If 1T Cu + Al Filter is not completely covered, adjust with
[SPM_CEST_XXXX]
[SPM_COLM_0007] PANO Collimator Position
[Pano Sensor Setting Command]