3. Remove the bite block and set the handle frame for PANO mode.
4. Press and hold the X-ray exposure switch when you see the message “HOLD THE BUTTON
(marked in the white box).
5. Release the switch when you see the message “RELEASE THE BUTTON” (marked in the
white box). If the system does not show this message, repeat step 3.
6-1. When you see a pop-up message notifying the end of auto PANO auto-alignment, click OK to
start the PANO collimator speed alignment.
6-2. When the system shows an error message, repeat steps 1 to 5 until the PANO auto-alignment
is completed.