EasyManua.ls Logo

Vatech A9 PHT-30CSS - Page 171

Vatech A9 PHT-30CSS
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 170 of 244
12. Click each data file in View 16. Ensure that the median value of each file meets the
standards below
Norm Bright standard value: 4 points with Cu 1 mm filter
Bright Min Max
Median value
Cal Point 1 200 300
Cal Point 2 300 500
Cal Point 3 500 800
Cal Point 4 800 1000
Fast Bright standard value: 4 points with Cu 1 mm filter
Bright Min Max
Median value
Cal Point 1 200 300
Cal Point 2 300 400
Cal Point 3 400 500
Cal Point 4 500 600
If the median values do not satisfy the standards, adjust the kVp0 and kVp1 values and
repeat the calibration procedure. (Adjustment range: ±10)
13. Move the CAL file that you previously saved OPEN folder to the CAL folder.
Norm C:\VCaptureSW\Acquisition\Ceph\WideCephG\Norm\CAL\OPEN
Fast C:\VCaptureSW\Acquisition\Ceph\WideCephG\Fast\CAL\OPEN
14. Check if 6 CAL files are saved in the CAL folder.

Table of Contents

Related product manuals