6. Set the Threshold value concerning the following standard, and check 0010.raw to check if it
satisfies the collimator alignment standard.
7. The collimator alignment standard is as follows.
Collimator Alignment standard value
Collimator Min pixels Max pixels
Threshold
90%
Left 1 5
Right 1 5
Upper 20 40
Lower 20 40
8. If the red pixel count is out of the permitted value, change the collimator position values by
[SPM_CAST_XXXXX].
Collimator Align can also be done using the EzAlign program provided by the
manufacturer.