6.3.2. CEPH Collimator Alignment with EzAlign
Follow the steps below to set CEPH collimator with EzAlign:
1. Start the EzAlign-S and click Modality Align.
2. Click CEPH and START. After clicking these buttons, the system moves to its capturing
position and the collimator auto-alignment starts.
If you want to use the auto irradiation function, click AUTO EXPOSE.
3. Remove the bite block and set the CEPH ear rods.