2.3.5. CT sensor: Bright Calibration
Stay outside of the X-ray room (imaging room) when the system is
collecting the bright calibration data.
2.3.5.1 How to Perform 2-Point Bright Calibration
To perform a 2-point bright calibration:
1. Run the C:\VCaptureSW\Acquisition\CBCT\WidePANO\VACAP.exe file.
2. Click the Settings tab and follow the step below.
1) Enter 2 for the Numbpoint.
2) Check if the tube voltage (kVp) and current (mA) meets the standards. If necessary, enter
section according to the table below:
Section Values
kVp0 630
mA0 700
kVp1 750
mA1 910
3) Go to the 4 section and insert “//” before all data (example: // kVp0=570)
4) Click Save.