Page 65 of 244
6. Click Bright.
7. When you see the message, “HOLD THE BUTTON FOR ~3 SEC” on the status window,
press the X-ray exposure switch.
8. Release the switch when the message, “RELEASE THE BUTTON” shows up in the status
window.
9. Repeat steps 7 to 8 unti you acquire two bright calibration data files.
10. Click V to open View16.