Equipment Safety Instructions
This electro-medical device contains high voltage and current flowed parts, and the following
instructions must be followed to avoid critical damage to the body and equipment.
Electrical Safety Caution
• The equipment and PC should be connected to a common
protective earth. Confirm that the equipment is well-
grounded.
• High voltage and current flows throw equipment. Turn off
equipment power and wait at least 2 minutes before the
operation begins.
• Do not connect this equipment to any device without
authorization from VATECH Co., Ltd.
• Do not connect this equipment to a Multiple Portable
Socket- Outlet (MPSO) or extension wires
• Keep in mind that MCU boards are electrically sensitive
that operators need to wear shielding gloves while
operating.
• The only part that can be replaced by the users is the input
fuses. Check if the fuses comply with the manufacturer’s
specifications.
Mechanical Safety Caution
• Be aware that fingers or parts of your body are not shut in
the equipment.
• If a problem occurs during operation, immediately press the
red emergency stop button to stop equipment operation
and cut off all power to the equipment.
• Never touch the patient while touching the connector of
SIP/SOP.
• While adjusting the patient, ensure that the laser beam is
not directed at the patient’s eyes.
• Never expose the equipment to liquids, mists, or sprays.
Any liquid may cause an electrical shock or otherwise
damage the system.
• Do not use spray cleaners on the equipment to avoid fire.
• Do not use solvent aerosol or spray cleaning agents
directly on the surface of the equipment. It can damage
equipment, including the LCD panel.