EasyManua.ls Logo

Vatech A9 PHT-30CSS - Page 70

Vatech A9 PHT-30CSS
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 69 of 244
9. Repeat steps 7 to 8 until you acquired four bright calibration data files.
10. Click the V to open View16.
11. Click each file in View 16 and check whether each median value satisfies the standards in
the table below:
Bright Min Max
Median
value
Cal Point 1 80 120
Cal Point 2 150 700
Cal Point 3 800 1800
Cal Point 4 2000 2500
[4-Point Bright Calibration Standard Value]
If the median value falls outside of the standard, adjust the kVp0 and kVp1 values and
repeat the calibration procedure. (Adjustment range:±10)
12. Copy the saved data from the OPEN folder and paste them to the CAL folder. After this step,
the CAL folder must contain 6 files.
Point 1
Point 2
Point 3
Point 4

Table of Contents

Related product manuals