VTK‐VS‐BA‐004e‐B|UserManualMDS3200A | Operation
37
6.8 LoadingtheFactorySettings
ByrecallingtheFactorySettings(implementedbythemanufacturer),youreturntoaprede‐
finedstartingposition.Thisisusefultostarttheinputofanewparameterset.
ThesearethevaluescontainedintheFactorySettings:
RI=0,50ms,FA=0.20ms,OT=0ms,
NL=90%,DL10.0msandNP=1
Step1: Press[recall].
Step2: [DownwardArrow]immediatelyopensthisparameterset.
Step3: [enter]theselection.
IMPORTANTNOTE!
Press[esc]toabort.
6.9 SwitchingOFFtheMDS3200A
Step1: Thecurrentdispensingcyclemustbecompleted,sothatthevalveisinhomeposition.
Step2: Lowerthesupplypressuretoatmosphericpressure(becauseininitialposition,the
valveisstillopened).Disengagethepneumaticsupply.Ifrequired,closethecar‐
tridgebyusinglockingpinMDT
309.
Step3: SwitchOFFthecontrolunit(ON/OFFbuttonattherearend).
Step4: Disconnectthevalvefromfluidsupply.
Step5: Removeallcablesfromthevalve.
Step6: Unscrewthescrewsusedforfixingthevalveinplace.
Afterworkingwithself‐curingsubstances,immediatecleaningis
advisable
inordertoavoidclogging.Thisconcernsthevalveitselfandallpartsincon‐
tactwiththefluid.Thevalveanditssubcomponentshavetobedismantled
andcleaned,asdescribedinthechapter“Cleaning”.
CAUTION!(preparecleaning)
Itisrecommendedtopreparethecleaningprocedureinadvance.
Donotdropthedeviceoritssubcomponents.