Message Possible cause Suggested action Alarm priority
applicable SpO2 tester. If the
message persists, replace the
module.
SpO2 initializing. 041600 None
None.
1
Very low
Searching for pulse
signal. 041800
SpO2 pulse search Pulse search is part of normal
operation, and does not have an
associated corrective action.
High
Low perfusion index.
Clear to retry. 041a00
There is marginal SpO2 pulse
quality or artifact.
Reapply the sensor to the patient. If
the problem is still present, replace
the SpO2 sensor. If the problem is
still present, replace the cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists
replace the module.
Very low
Demo Mode active.
Restart the device.
041b00
The SpO2 parameter is in demo
mode
None.
2
Very low
Sensor not connected.
Clear to retry. 041c00
Check the SpO2 sensor connection Check the sensor and cable
connection. If the problem is still
present, replace the SpO2 sensor. If
the problem is still present, replace
the cable. If the problem is still
present, verify the module
functionality by replacing the sensor
with an applicable SpO2 tester. If the
message persists replace the
module.
Very low
SpO2 not functional.
041d00
There is an SpO2 raw queue
overrun
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 04ef00 There is an SpO2 raw queue
overrun
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
170
Appendix Connex
®
Spot Monitor