BOD
149
9BOD
For the inoLab Level 3 BSB/BOD measuring module.
9.1 General information
For BOD measurements, the inoLab BSB/BOD enables:
l The administration of a maximum of 30 samples.
l The use of a maximum of 18 sample dilutions per sample.
l The use of a maximum of 5 blank value solutions (for
determination of the blank value).
l The setting of the incubation time (5 to 30 days).
l The assignment of names to the blank value solutions,
samples, sample dilutions and routines (max. 12
characters for each name and max. 7 routines).
l The simplification of entries by the use of an external IBM-
PC keyboard (see section 4.2.3).
l Automation by the use of BOD routines (see section 9.4);
This enables BOD measurements to be automated by
pressing a key for each measurement.
l Automation by using the StirrOx G oxygen sensor with
integrated stirrer;
This enables measurement routines to be carried out
simply with the key on the sensor.
l Automation through the use of a barcode reader (see
section 9.1.2);
This makes the entry of the names of samples, sample
dilutions and blank value solutions unnecessary.
l The editing and printout of the measurement results (see
section 9.3.1).