What to do if...
237
14 What to do if...
System messages due to errors are listed together with their
possible causes and remedies.
14.1 Terminal messages
Probe symbol
flashes
Instrument does not
react to keystroke
h flashes
continuously
Cause Remedy
– Calibration interval
expired
– Recalibrate the measuring
system
Cause Remedy
– Operating state
undefined or
electromagnetic
interference too high
– Processor reset:
Unplug the plug-in power
supply and plug it in again
(if it occurs repeatedly,
ensure the
electromagnetic radiation
is reduced)
Cause Remedy
No stable measured value Provide stable measuring
conditions (e.g. temperature)
Defective probe Use another probe