14.5 Oxygen system messages
Measuring modules for the inoLab Level 3:
pH, BSB/BOD, pH/Oxi and Multi.
Error message
OFL
Error message
E3
Error message
E7
Cause Remedy
Display range exceeded
Not connected Connect the probe
Cable broken Replace probe
Depleted Replace probe
Short-circuit between gold
and lead electrode
Clean probe and
replace it if necessary
Cause Remedy
Invalid calibration
Electrolyte solution
depleted
Regenerate probe
Membrane contaminated Clean membrane
Electrode system poisoned Regenerate probe
Worn out Replace probe
broken Replace probe
Cause Remedy
Membrane damaged
Membrane damaged Regenerate probe
Membrane head not
screwed on tight enough
Screw membrane head
tight