239
14.3 pH/voltage system messages
Measuring modules for the inoLab Level 3:
pH, pH/ION, pH/Cond, pH/Oxi and Multi.
Error message
OFL
Error message
E3
Cause Remedy
pH/voltage, electrode:
Not connected Connect electrode
Air bubble in front of the
diaphragm
Remove air bubble
Air in the diaphragm Extract air or
moisten diaphragm
Cable broken Replace electrode
Gel electrolyte dried out Replace electrode
Cause Remedy
pH/ORP electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Moisture in the plug Dry plug
Electrolyte out of date Replenish electrolyte or
replace electrode
Electrode worn out Replace electrode
pH electrode: broken Replace electrode
ORP electrode: short-
circuit
Replace electrode