EasyManua.ls Logo

wtw Multi 3620 IDS - Page 35

wtw Multi 3620 IDS
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
pH value Multi 3620 IDS
35 ba77169e01 04/2016
1 In the measured value display, select the measured parameter pH or
mV with <M>.
2 Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
3 Thoroughly rinse the IDS sensor with deionized water.
4 Immerse the IDS sensor in buffer solution 1.
5 For measurements without temperature sensor
(e.g. when using an IDS adapter
):
Measure the temperature of the buffer manually and enter it with
<><>.
6 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
7 Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.
15.03.2016 08:00
ConCal
15.03.2016 08:00
ConCal

Table of Contents

Related product manuals