Multi 3620 IDS Dissolved oxygen
ba77169e01 04/2016 54
7.3.5 Calibration data
The calibration data can be displayed and then output to the interface.
Displaying the
calibration record
The calibration protocol of the last calibration is available in the menu Calibra-
tion / Calibration record. To open it in the measured value display, press the
<CAL_> key.
The calibration records of the last 10 calibrations are available in the menu Cal-
ibration / Calibration data storage / Display. To open the Calibration menu in
the measured value display, press the <ENTER> key.
Calibration
evaluation
After calibrating, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item Setting/func-
tion
Explanation
Calibration /
Calibration data stor-
age / Display
- Displays the calibration records.
Further options:
Scroll through the calibration
records with <><>.
Output the displayed calibra-
tion record to the interface with
.
Output all calibration records
to the interface with <PRT__>.
Quit the display with <[ESC]>
or <ENTER>.
Switch directly to the mea-
sured value display with <M>.
Calibration /
Calibration data stor-
age / Output to USB
flash drive or printer
- Outputs the stored calibration
data to the USB-A interface (USB
memory device/USB printer)
Calibration /
Calibration data stor-
age / Output to
RS232/USB
- Outputs the stored calibration
data to the USB-B interface (PC)