Multi 3620 IDS Dissolved oxygen
ba77169e01 04/2016 50
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
7.1.2 Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.
7.2 FDO
®
check procedure (check of the FDO
®
925)
7.2.1 Why should you check the sensor?
With the FDO
®
Check procedure, you can find out in a simple manner whether
the FDO
®
925® D.O. sensor should be cleaned or calibrated.
7.2.2 When should you check the sensor?
Checking can be useful in the following cases:
When the check interval has expired
If the measured values seem to be implausible
If you assume that the sensor cap is contaminated or at the end of
its lifetime
After the sensor cap was exchanged
3 Using <ENTER>, start a further measurement with Stability control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Measured
parameter
Time interval Stability in the time inter-
val
D.O. concentra-
tion
20 seconds
: better than 0.03 mg/l
D.O. saturation 20 seconds
: better than 0.4 %
D.O. partial
pressure
20 seconds
: Better than 0.8 mbar
Temperature 15 seconds
: better than 0.5 °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3620 IDS and is the answer not in the manual?

wtw Multi 3620 IDS Specifications

General IconGeneral
Brandwtw
ModelMulti 3620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals