Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values. The display of the measured parameter flashes until
a stable measured value is available.
You can start the Stability control function manually at any time, irrespective of
the setting for automatic Stability control (see section 11.6.3) in the System
menu.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
2 Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.
1 Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2 Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the dis-
play of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3 Using <ENTER>, start a further measurement with Stability control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.