7-8 <7. Trouble Shooting>
IM 80J01A01-01E
Problem
Probable Cause / Corrective Action
It is impossible to acquire time-lapse
by set interval
It might be that time for acquiring
sequence per one time point exceeds
time-lapse interval.
󳊜 Set interval longer or reduce
number of well, field and Z slice so
that acquiring sequence per one
time point doesn’t exceed interval
time
Condensation occurs on the rid of
sample vessel when using Stage
Heater
It might be caused by dust or water
on the condensation reducer.
󳊜 Clean the condensation reducer
by referring 7.2.1 Cleaning
Procedure of the Condensation
Reducer
Condensation occurs on room
temperature sample when using
Stage Heater.
󳊜 Please warm the sample up to
30 󳇾 and over.
󳆽
1
In the case of inquiring the dealer about trouble, it may need to send system log file.
For outputting of log file, please refer to page 5-7.
7.2.1 Cleaning Procedure of the Condensation Reducer
Click Sample tab “OPEN” Button of CQ1 Software or do long press
(3 seconds) of Door OPEN/CLOSE Button of CQ1 Main Unit to open
the door.
Door OPEN/CLOSE Button