For the 701255 (NONISO_10M12)
This module is non-isolated. Be sure to fasten the module screws when measuring a
voltage above 42 V on this module. In addition, use the 701940 (10:1) dedicated non-
isolated passive probe.
Maximum input voltage (at a frequency of 1 kHz or less)
Combined with the 701940 (10:1) passive probe
7
600 V (DC + ACpeak)
Direct input (cable that does not comply with the safety standards)
5
200 V (DC + ACpeak) (as a value that meets the safety standard)
250 V (DC + ACpeak) ( maximum allowable voltage, as a value that does not damage
the instrument when applied)
For the 701275 (ACCL/VOLT)
Maximum input voltage (at a frequency of 1 kHz or less)
Combined with the 701940 (10:1) passive probe
7
or direct input (cable that does not
comply with the safety standards)
5
42 V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
Combined with the 701940 (10:1) passive probe
8
or direct input (cable that does not
comply with the safety standards)
6
42 V (DC + ACpeak) (CAT II, 30 Vrms)
For the 701281 (FREQ)
Maximum input voltage (at a frequency of 1 kHz or less)
In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a passive
probe (10:1) 702902
1
420 V (DC + ACpeak)
Safety cable (1:1) (combined with 701901+701954)
3
or direct input (cable that does not
comply with the safety standards)
5
42 V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a passive
probe (10:1) 702902
2
300 Vrms (CAT II)
Safety cable (1:1) (combined with 701901+701954)
4
or direct input (cable that does not
comply with the safety standards)
6
42 V (DC + ACpeak) (CAT II, 30 Vrms)
For the 720243 (SENT)
Maximum input voltage (at a frequency of 1 kHz or less)
In combination with an isolated probe (10:1) 700929 or a passive probe (10:1) 702902
1
420 V (DC + ACpeak)
Safety cable (1:1) (combined with 701901+701954)
3
or direct input (cable that does not
comply with the safety standards)
5
42 V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
In combination with an isolated probe (10:1) 700929 or a passive probe (10:1) 702902
2
300 Vrms (CAT II)
Safety cable (1:1) (combined with 701901+701954)
4
or direct input (cable that does not
comply with the safety standards)
6
42 V (DC + ACpeak) (CAT II, 30 Vrms)
3.5 Connecting Probes
3-19
IM DL850E-03EN
Making Preparations for Measurements
1
2
3
4
5
6
App
Index

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL850E and is the answer not in the manual?

YOKOGAWA DL850E Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL850E
CategoryTest Equipment
LanguageEnglish

Related product manuals