Specifications of the Probe Power Supply Terminals (Optional)
Item
Number of probes that can be used 4
Probes that can be used Current probe (701930, 701931, 701932, and 701933)
Number of current probes that can be used 701930 (150 A): Up to 2
701931 (500 A): Up to 1
701932 (30 A): Up to 2
701933 (30 A): Up to 2
Supply voltage ±12 V, four outputs (up to 1 A total current)
* Up to four 700937 current probes can be used, but this probe is a discontinued product.
When using the current probe, the number of probes that can be used is limited by the current
generated by the device under measurement (current measured by the current probe). The measured
versus consumed current characteristics of current probes that can be connected to the instrument are
indicated below.
-250
-200
-150
-100
-50
-15 -10 -5 0 5 10 15
0
50
100
150
200
250
Current probe (700937)
Current consumption (mA)
Measured current (A)
Positive
current
Negative
current
-400
-300
-200
-100
-150 -100 -50 0 50 100 150
0
100
200
300
400
Current consumption (mA)
Measured current (A)
Positive
current
current
AC (f = 50 Hz)
AC (f = 50 Hz)
-500
-400
-300
-200
-100
-500 -400 -300 0 100 200 500
0
100
200
300
400
500
Measured current (A)
Positive
current
Negative
current
-200 -100 300 400
Current probe (701931)
AC (f = 50 Hz)
DC
Current probe (701932)
Current consumption (mA)
Measured current (A)
600
500
400
300
200
100
0
-100
-200
-300
-400
-500
-600
Positive
current
current
AC (f = 50 Hz)
DC
-50 -40 -30 -20 -10 0 10 20 30 40 50
Current probe (701933)
Current consumption (mA)
Measured current (A)
600
500
400
300
200
100
0
-100
-200
-300
-400
-500
-600
Positive
current
Negative
current
AC (f = 50 Hz)
DC
-50 -40 -30 -20 -10 0 10 20 30 40 50
For details about each current probe, contact your nearest YOKOGAWA dealer.
3.5 Connecting Probes