EasyManuals Logo

YOKOGAWA WT3002E User Manual

YOKOGAWA WT3002E
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
1-3
IM WT3001E-51EN
Motor Evaluation Function (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
Torque Signal Input Connector (TORQUE)
Input the signal output from the torque meter (a DC voltage (analog or pulse signal) that
is proportional to the torque of the motor) according to the following specifications.
TORQUE
20V MAX
DC Voltage (Analog Input)
Item Specifications
Connector type
Input range
Effective input range
Input resistance
Maximum allowable input
Continuous maximum common mode voltage
Isolated BNC connector
1 V, 2 V, 5 V, 10 V, and 20 V
0% to ±110% of the measurement range
Approx.1MΩ
±22 V
±42 Vpeak or less
Pulse Input
Item Specifications
Connector type
Frequency range
Amplitude input range
Effective amplitude
Input waveform
Input resistance
Continuous maximum common mode voltage
Isolated BNC connector
2 Hz to 200 kHz
±12 Vpeak
1 V (peak to peak) or more
50% duty ratio rectangular wave
Approx.1MΩ
±42 Vpeak or less
1.2 Applying Signals of Rotating Speed and Torque