7.8  Setting the Anti-Aliasing Filter
Procedure
 
START  STOP 
REMOTE 
CAL 
MOTOR 
SET 
CURSOR  NULL 
MENU  STORE SET 
LOWER ITEM  LOWER FORM  ALL USER SET 
RESET  SET 
PAGE  PAGE 
SCALING 
HRM SET 
MEASURE 
SYNC SOURCE 
AVG 
LINE FILTER 
FREQ 
FILTER 
 
 
FILE 
PRINT 
MISC 
STORE 
IMAGE SAVE 
HELP 
NUMERIC 
WAVE 
OTHERS 
ITEM 
FORM 
U / I / P 
WP/q/ 
TIME 
USER 
ELEMENT 
UPDATE 
RATE 
HOLD 
SINGLE 
INTEG 
LOCAL 
SHIFT 
Cursor keys 
To exit the menu during operation, press ESC located above the 
soft keys.
In the procedural explanation below, the phrase “press the cursor 
keys” may be used.  This phrase refers to the procedures for 
selecting items and entering values and character strings.  For 
details on the procedures, see section 3.14 in the User’s Manual, 
IM WT3001E-01EN.
When making harmonic measurements, the line filter of normal measurement mode is 
used as an anti-aliasing filter.
•  Selecting the Line Filter
1.
 Press LINE FILTER to display the Line Filter menu.
  Only the installed elements are displayed.
•  Selecting the Target Element
2.
  Press one of the soft keys corresponding to the displayed element to select the 
target element.
•  Selecting the Line Filter
3.
  Press the cursor keys to select any filter between OFF and 50 kHz. If any of 
the target elements is set to a setting other than OFF, the LINE FILTER key 
illuminates.