8-8
IM WT3001E-51EN
Frequency Measurement
Item Specifications
Measurement range • PLLsynchronizationmethod:2.5Hz≤f≤100kHz
• Externalsamplingclockmethod:0.15Hz≤f≤5kHz
Measurement Function
Item Specifications
Display update Depends on the PLL source
• PLL synchronization method: 1 s or more
• External sampling clock method: 20 s or more
PLL Lock Timeout Warning
Item Specifications
Timeout value Depends on the PLL source
• PLL synchronization method: 5 s or more
• External sampling clock method: 40 s or more
Functional Limitations
Item Specifications
Display Waveform display, trend display, computed waveform display, and FFT display are not allowed.
Averaging Moving average is not allowed.
Data update interval Depends on the PLL source
Voltage/Current mode Not selectable. The LED is OFF.
Peak measurement Not allowed
Efficiency computation Not allowed
Compensation Efficiency compensation is invalid.
Integration Not allowed
Store Integration synchronization (Integ Sync) is not allowed. Waveform storage is not allowed.
File save Waveform and waveform sampling data (Acquisition) are not allowed
NULL Not allowed
Synchronized measurement Not allowed
Motor evaluation function Not allowed
Print Integration synchronization using auto print is not allowed.
Delta computation Not allowed
Measurement Functions
For a description of the measurement functions of wide bandwidth harmonic
measurement mode and how they are determined, see section 7.11. Measurement is not
allowed for measurement functions related to the functions above that have limitations.
For a list of measurement functions that cannot be measured, see appendices 10 and 11
in the User’s Manual IM WT3001E-01EN.
8.5 Wide Bandwidth Harmonic Measurement Specifications