9-11
IM WT3001E-51EN
IEC Harmonic Measurement (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
Item Specifications
Accuracy When the line filter (5.5 kHz) is ON
Frequency Voltage and Current
±(reading error + measurement
range error)
Power
±(reading error + measurement
range error)
45Hz≤f≤66Hz
66Hz<f≤440Hz
440Hz<f≤1kHz
1kHz<f≤2.5kHz
2.5kHz<f≤3.3kHz
0.2% of reading + 0.04% of range
0.5% of reading + 0.05% of range
1% of reading + 0.05% of range
2.5% of reading + 0.05% of range
8% of reading + 0.05% of range
0.4% of reading + 0.05% of range
1.2% of reading + 0.1% of range
2% of reading + 0.1% of range
5% of reading + 0.15% of range
16% of reading + 0.15% of range
However, all the items below apply.
When the crest factor is set to 3
• Whenλ(powerfactor)=1
Power figures that exceed 440 Hz are reference values.
For external current sensor range, add 0.03 mV to the current accuracy and add (0.03 mV/
externalcurrentsensorrangerating)×100%ofrangetothepoweraccuracy.
For direct current input range on the 30-A input element, add (0.1 mA/direct current input
rangerating)×100%ofrangetothepoweraccuracy.
For direct current input range on the 2-A input element, add (1 µA/direct current input range
rating)×100%ofrangetothepoweraccuracy.
For direct current input in a range less than or equal to 200 mA on the 2-A input element, add
0.02%ofreading+0.01%ofrangetothecurrentaccuracyintherangeof45Hz≤f≤66Hz
and add 0.03% of reading + 0.01% of range to the power accuracy.
For n
th
order component input, add ({n/(m+1)}/50)% of (the n
th
order reading) to the n+m
th
order and n-m
th
order of the voltage and current, and add ({n/(m+1)}/25)% of (the n
th
order
reading) to the n+m
th
order and n-m
th
order of the power (only when applying a single
frequency).
Accuracy when the crest factor is 6: The same as when the range is doubled for crest factor 3.
The accuracy guaranteed range by frequency and voltage/current is the same as the
guaranteed range of normal measurement.
9.6 IEC Harmonic Measurement Specifications